Choosing & Preparing Biological Probes for FCS Assays

High quality fluorescently labeled probes are required for successful fluroescence correlation spectroscopy measurements. The following lists describe key probe properties to consider as you are designing FCS experiments.

Probe-Target Considerations

  • Affinity of Probe for Target: It is critical to understand the affinity of probe-target interactions in any FCS assay being conducted (See Biological Interactions that FCS Can Detect). Low affinity probes require high concentrations, while high affinity probes allow you to use low concentrations.
  • Selectivity of Binding: High sensitivity measurements require highly selective probes that only bind to the intended target. If there is non-specific binding, or even specific binding to small quantities of intefering substances(e.g. buffer components, carrier proteins such as BSA, sample chamber) in the sample, FCS results may be ambiguous.

Probe-Dye Considerations

  • Purity of the Probe: Probe purity will have a large impact on the quality of your FCS measurements. Probe samples should contain as little free dye as possible (See Choosing Fluorescent Dyes for FCS Assays). HPLC purification and a variety of spin purification schemes exist to ensure > 95% purity. Correlation analysis will easily detect free dye in excess of 5-10% of probe fluorescence. Although it is still possible to quantify FCS measurements made under these conditions, it does reduce the useful range over which assays can be conducted and may require more complex data analysis.
  • Specific Activity (dyes per molecule) of the Probe: The total number of dye molecules per labeled probe is the Specific Activity or Stoichiometry of labeled probes. This specific activity can be modulated to extend the useful range over which you can conduct FCS measurements using the QuantumXpert. In addition, knowing the specific activity enables you to quantify binding stoichiometry.
  • Affect of Dye Labeling on Probe Affinity: It is important to understand how labeling a probe affects binding affinity, especially when using it is used in competition binding experiments. If the biological activity of the probe is significantly altered by conjugation to a fluorescent dye (e.g. reduced affinity), it may not be ideal for use in highly sensitive FCS measurements.
  • Affect of Probe-Target Binding on Dye Quantum Yield: If quantum yield is altered when a labeled probe binds to its target, or when large complexes form, this must be accounted for when analyzing FCS data.

Choosing Fluorescent Dyes for FCS Assays

When selecting a fluorescent dye to use in a fluorescence correlation spectroscopy assay, the first consideration should be compatibility with the biological or chemical system to be studied.

Ideal FCS Dye Properties:

  • Little or no nonspecific binding to your probe or target.
  • Resistant to irreversible photobleaching.
  • Resistant to triplet state excitation
  • Large emission at low intensity excitation (e.g. high quantum yield)

The dyes listed below have been used successfully in FCS Assays, and are compatible with the QuantumXpert FCS Spectrometer. Although we have not yet compiled a comprehensive characterization of all of these dyes, we are working on it!

Fluorescent Dyes for Covalent Conjugation

  • Rhodamine 6G
  • Alexa Dyes (488, 532, 568)
  • Cy 3.5 and Cy 3.5b
  • Tetramethyl Rhodamine
  • Texas Red
  • Quantum Dots (525, 565, 605)

Intrinsically Fluorescent Proteins

  • GFP
  • CFP
  • YFP

Dyes That Selectively Bind or Intercalate

  • DiI

Choosing Buffers for FCS Assays

Although it is of primary importance to select buffers compatible with the biological and chemical interactions to be studied, you must also meet requirements for effective FCS measurements.

Important Buffer Properties to Consider

pH: In addition to affecting the interactions and stability or your biological probes, buffer pH can affect the function of many fluorescent dyes

Salt Concentration: Salt concentrations can affect the stability and aggregation state of your probes. Additionally, high salt concentrations can be detrimental to some fluorescent dyes, such as Quantum Dots.

Detergents: Detergents can limit the amount of aggregation of your probes and reduce adsorption to the chamber, but can also affect fluorescent signal intensity and diffusion rates.

BSA: Adding BSA to your buffer can stabilize your biological probes, but may interfere with interactions or affect diffusion rates.

Other Additives: Any other additive must be measured to ensure it does not contribute significant background fluorescence to your sample.

List of Buffers Used in Published FCS Assays

  • PBS + 0.05% v/v Tween-20 (Casein cleavage and IL-4 receptor)
  • 100 mM diethanolamine-HCl (pH 10) and 1 mM MgCl2 (alkaline phosphatase hydrolysis)
  • 50 mM Tris-HCl (pH 7.6), 150 mM NaCl, 10 mM MgCl2, 10 mM DTT and NH2-sulfate (p21 ras nucleotide exchange)
  • 20 mM HEPES (pH 7.4), 50 mM NH4Cl, 0.05% Tween20, 10 mM MgCl2 (RNA-ligand binding assay)
  • 50 mM Tris-HCl (pH 7.2), 10 mM MgCl2, 0.05% Tween20 (tyrosine kinase assay)
  • 50 mM MOPS (pH 7.5), 1 mM EDTA and 0.1% TX100 (trypsin and chymotrypsin assays)
  • 600 nM substrate, 1-2 uM enzyme in 50 mM HEPES (pH 7.5), 30% sucrose, 150 mM NaCl, 0.1% Tween20, 0.01% PEG3350 (CMV protease assay)
  • 50 mM MOPS (pH 7.5), 1 mM EDTA, 0.1% TX-100 with reaction termination by dilution 100-fold with 300 nM avidin in same buffer (leader peptidase assay)
  • 140 mM NaCl, 20 mM HEPES (pH 7.4), 5 mM MgCl2, 1.8 mM CaCl2, 0.36 g/L NaHCO3, 1 g/L D-Glucose (EGF binding to EGF-R in vesicles)
  • 60 mM KCl, 5 mM MgCl2, 10 mM HEPES (pH 7.9), 250 mM ZnSO4, 2 ug/reaction poly-DC with 6.8 nM oligonucleotide and 0.3 to 3 ul nuclear extract (protein-DNA interaction…transcription factor)
  • 60 mM KCl, 5 mM MgCl2, 10 mM HEPES pH 7.5 (thyroid hormone receptor/DNA interaction)
  • 0.18M NaCl (DNA hybridization studies)

Choosing Component Concentrations for FCS Assay

When choosing the concentration of each of the components in your FCS assay sample, the following properties should be considered.

Probe Affinity: Probe and target concentrations should be in the optimum range at which biological or chemical interactions occur.

Dye Concentration: The QuantumXpert FCS Spectrometer is intended to measure fluorescent dye concentrations from ~50 pM to ~50nM. The number of dye molecules per probe times the concentration of the probe should fall within these boundaries. Specific activity (dyes per probe) can be adjusted so that the probe concentration can be well outside of the 50pM - 50nM boundaries.

Aggregation Susceptibility: Many probes, and also fluorescent dyes, will aggregate at high concentrations.

Advantages of AssayXpert Kits for FCS Assays

AssayXpert Kits are developed for use with QuantumXpert FCS Spectrometers.

The following properties make AssayXpert Kits ideal for further simpifying your fluorescence correlation spectrscopy experiments.

Selective Probes: AssayXpert Kits have been designed with highly selective probes and are characterized to quantify the contribution of non-specific binding

Purified Probes: AssayXpert Kits contain probes that have been purified. Free dye typically makes up <5% of the labeled probe.

Stable Dyes: AssayXpert Kits utilize fluorescent dyes that are resistant to irreversible photobleaching and intersystem crossing.

Compatible Dyes: AssayXpert Kits utilize fluorescent dyes that have been thoroughly characterized in the QuantumXpert. Crosstalk between emission channels is automatically corrected using the included software analysis module.

Optimized Conditions: Each AssayXpert Kit has been optimized to detect the analyte of interest and buffer components are carefully controlled to eliminate sample artifacts. AssayXpert Kits come with all reagents necessary to detect the unlabeled analyte of interest and requires no fluorescent labeling by the end-user.

Customized Software: Each AssayXpert Kit comes with a software analysis module with a simplified, assay-specific interface.